DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Gas/Air Leak (2946)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.And found that the helium reservoir, high pressure regulator, regulator housing and helium tubing were all leaking.The cable for the helium regulator was corroded also.The fse replaced the 9 volt battery, helium reservoir, helium regulator, high pressure regulator, regulator housing, helium tank transducer cable, and helium tubing.The fse performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a helium leak.It is unknown the circumstances under which the event occurred.However, there was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a helium leak.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involved, however, no patient harm or adverse event was reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, b6, b7, g3, g4, g7, g8, h2, h6(investigation type, component codes & impact codes), h10, h11.Corrected fields: b5, d4(model#), d5, d11, g1(contact person), h6(clinical code).
|
|