This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p), implanted with another manufacturer's right ventricular (rv) lead, triggered a lead safety switch (lss) due to high out-of-range pace impedance measurements greater than 3000 ohms.Now in a unipolar configuration, there was noise noted due to myopotential with greater than two seconds of pacing inhibition.Ts reviewed troubleshooting and programming options.The patient was seen in clinic, however noise was not reproducible.Review of stored episodes revealed inappropriate signal artifact monitoring (sam) episodes attributed to atrial fibrillation (af).Technical services (ts) discussed considering vvir to turn off right atrial (ra) sensing for the inappropriate sam episodes.For the lss, the plan was to leave the device in unipolar pacing and bipolar sensing and decrease both rv and left ventricular (lv) sensitivity to 5 mv.The patient will continue to be monitored.This device remains in service.No adverse patient effects were reported.
|