• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACCOLADE MRI EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L331
Device Problems Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438)
Patient Problems Arrhythmia (1721); Palpitations (2467)
Event Date 12/28/2020
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this patient presented with palpitations.Device interrogation identified the lead safety switch had been triggered due to an out of range atrial pacing impedance measurement.A review of the presenting data identified noise, oversensing and suspected loss of capture.A boston scientific technical services consultant discussed the clinical observations with the caller who stated the patient will undergo additional evaluation.No adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.
 
Event Description
This supplemental report is being filed to update the investigation conclusion code and additional manufacturing narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE MRI EL DR
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key11233353
MDR Text Key228849145
Report Number2124215-2021-02127
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526559266
UDI-Public00802526559266
Combination Product (y/n)N
PMA/PMN Number
P150012/S000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/12/2022
Device Model NumberL331
Device Catalogue NumberL331
Device Lot Number875067
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/29/2020
Initial Date FDA Received01/26/2021
Supplement Dates Manufacturer Received02/20/2021
Supplement Dates FDA Received04/16/2021
Patient Sequence Number1
Patient Age48 YR
-
-