Model Number L331 |
Device Problems
Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438)
|
Patient Problems
Arrhythmia (1721); Palpitations (2467)
|
Event Date 12/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that this patient presented with palpitations.Device interrogation identified the lead safety switch had been triggered due to an out of range atrial pacing impedance measurement.A review of the presenting data identified noise, oversensing and suspected loss of capture.A boston scientific technical services consultant discussed the clinical observations with the caller who stated the patient will undergo additional evaluation.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
This supplemental report is being filed to update the investigation conclusion code and additional manufacturing narrative.
|
|