| Model Number |
V172 |
| Medical Device Problem Codes |
Over-Sensing (1438); Pacing Problem (1439); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Syncope (1610); No Code Available (3191); Syncope/Fainting (4411)
|
| Date of Event |
01/03/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This device was not returned as our records indicate it was retained by the explanting hospital.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had entered safety mode, and the associated codes were 0x40, 0x0, 0x0.The patient presented to the emergency room (er) due to syncope.Additional information received indicated that this crt-p exhibited oversensing and pacing inhibition with greater than two seconds of asystole.This crt-p was explanted.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had entered safety mode, and the associated codes were 0x40, 0x0, 0x0.The patient presented to the emergency room (er) due to syncope.Additional information received indicated that this crt-p exhibited oversensing and pacing inhibition with greater than two seconds of asystole.This crt-p was explanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This device was not returned as our records indicate it was retained by the explanting hospital.If information is provided in the future, a supplemental report will be issued.
|
| |
|