| Model Number |
7311 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
Loss of consciousness (2418)
|
| Date of Event |
01/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter , so personal information was not entered.No information was captured in section as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
An advocate called on behalf of the customer to report that they obtained blood glucose readings of 245 mg/dl at 11:26 a.M.On the contour next ez meter and 121 mg/dl at 11:28 a.M.On the contour next meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour next ez meter and contour next test strips from lot # 0epeg04b for evaluation.There were only two strips remaining in the vial of the returned test strips.Therefore, the in-house contour next test strips from lot # 0epeg04b were also used for testing.The suspected contour next meter was not returned for evaluation.Therefore, an in-house contour next meter was tested with a returned test strip and in-house test strips using a blood sample, which gave a satisfactory performance.Additionally, the returned meter was tested with a returned test strip and in-house test strips using a blood sample, which gave a satisfactory performance.
|
| |
|
Search Alerts/Recalls
|