| Model Number |
IPN913861 |
| Medical Device Problem Code |
False Alarm (1013)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported the "respiratory therapist noticed that the neptune heater's cautionary and low water notifications were alarming.The water cube was replaced and the heater still alarmed.The rt then replaced the concha column and the alarms stopped".No patient injury reported.Patient condition reported as "fine".
|
| |
|