MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050 USA#HLS SET ADVANCED 7.0
Device Problem Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Date 04/09/2018
Event Type  malfunction  
Manufacturer Narrative
In regards to the issue concerning the punctures from inside to outside an inappropriate handling (upside down) can be concluded.If the welding points of the inlay become loose, it will allow the product within the inlay to move.As a consequence, the holder of the venous probe may cause damages in terms of two punctures through the tyvek cover.Based on this the failure could be confirmed.Furthermore a dhr review was conducted for the lot in question.There was no rework or scrap record performed during production activities.A trend review was performed.156 similar complaints were found.Due to this no systemic issue could be determined.The failure was detected during incoming inspection.The product was not used for patient treatment.The event has been reported with a delay due to our retrospective examination of the record.At the time (2018-04-09) the complaint was reviewed and found not to be reportable.Current day, we compared the record with equivalent events, and found an inconsistency with the reporting decisions.With the current knowledge and the current team of complaint handlers, we have come to conclude the event should have been reported.As a remedial effort, we will report it within capa # (b)(4).
 
Event Description
Customer called to claim that qty.2 hls 7.0 (part # 701052794) were received damaged with punctures on top.Complaint id (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLS SET ADVANCED
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
NURSEL BOELENS
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
4972229321
MDR Report Key11236424
MDR Text Key229389831
Report Number8010762-2021-00078
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
PMA/PMN Number
K112360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date02/22/2020
Device Model NumberBEQ-HLS 7050 USA#HLS SET ADVANCED 7.0
Device Catalogue Number701052794
Device Lot Number70123179
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/09/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/23/2018
Type of Device Usage Initial
Patient Sequence Number1
-
-