| Model Number |
1012449-06 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents and/or complaints from this lot.The investigation determined the reported complaint appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.Na.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a moderately tortuous and heavily calcified lesion in the left anterior descending artery.The balloon of a 3.0x6mm nc trek rx balloon dilatation catheter (bdc) ruptured during the first inflation at 12 atmospheres.The procedure was successfully completed with a 3.0x6mm non-abbott bdc.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|