| Model Number |
1896-5040S |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
01/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device remains in the patient.
|
| |
|
Event or Problem Description
|
|
As reported: "surgeon inserted a 5.0 locking screw acting as a "blocking screw." while inserting the scn (supracondylar nailing system), the nail broke the blocking screw in half.The surgeon was able to remove the top half of the screw, but the bottom half was left in the patient permanently.The scn was implanted and the rest of the case was completed successfully.".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: refer to h3 (reason code); h6 conclusion code.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However, based on the received information it¿s clear that locking screw was inserted prior to the scn nail to guide the nail as a blocking technique which is not specified in the respective operative technique of scn nailing system.Hence, it can be concluded that the event is user related (off label use).If device is returned or any further information is provided, the investigation report will be reassessed.H3 other text : device is not available; disposed by hospital.
|
| |
|
Event or Problem Description
|
|
As reported: "surgeon inserted a 5.0 locking screw acting as a "blocking screw." while inserting the scn (supracondylar nailing system), the nail broke the blocking screw in half.The surgeon was able to remove the top half of the screw, but the bottom half was left in the patient permanently.The scn was implanted and the rest of the case was completed successfully.".
|
| |
|