STERIS MEXICO, S. DE R.L. DE C.V. EVOLUTION TRANSFER CARRIAGE Back to Search Results
Device Problem Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/30/2020
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event.After speaking with the user facility, the technician was informed that the employee did not ensure the transfer carriage wheels were properly locked into the docking station when removing a load from the sterilizer.The technician found that the transfer carriage required adjustment due to damage as a result of the reported event.The technician made the necessary adjustments to the transfer carriage, tested the unit, confirmed it to be operating properly, and returned it to service.The technician counseled facility personnel on the proper use and operation of the transfer carriage and docking station, specifically, ensuring the wheels are properly locked to the docking station.No additional issues have been reported.
 
Event Description
The user facility reported that an employee was unloading a processed load from the sterilizer with their evolution transfer carriage when the carriage became unstable and fell to the floor.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION TRANSFER CARRIAGE
Type of Device
TRANSFER CARRIAGE
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key11240334
MDR Text Key238401725
Report Number3005899764-2021-00003
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/30/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-