SMITH & NEPHEW, INC. JRNY II TKA FEM SIZING GDE LT; PROSTHESIS, KNEE, PATELLOFEMTIB, SEMI-CONSTRAIND, CEMENTD, POLY/METAL/POLY
|
Back to Search Results |
|
| Model Number |
74012455 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the journey ii tka fem sizing gde left is not staying on 3 degree mark.Problem was noticed during a case.No injury to the patient or surgical delay was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device, used in treatment, was returned and evaluated.A visual inspection shows signs of wear and tear.A functional evaluation of the returned device confirmed the stated failure mode.The size adjuster on the device is damaged and loose, rendering it inoperable.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|