• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OPTIFLOW + MASK INTERFACE ADAPTER; BZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OPTIFLOW + MASK INTERFACE ADAPTER; BZA Back to Search Results
Model Number OPT980E
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).One of the complaint opt980 mask interface adapters is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of our investigation.
 
Event or Problem Description
A distributor in (b)(4) reported, via a fisher & paykel healthcare (f&p) representative, that opt980 mask interface adapters were "coming apart at the base of the tubing" during use.There were no patient consequences.
 
Event or Problem Description
A distributor in canada reported, via a fisher & paykel healthcare (f&p) representative, that three opt980e optiflow + mask interface adapters were "coming apart at the base of the tubing" during use.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(6) section d4 lot#: the lot number for one of the three complaint devices was provided and included in this report.The opt980e optiflow + mask interface adapter is used to deliver humidified respiratory gases to patients.The interface is held in place by a head strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's interface.Method: one of the three complaint opt980e optiflow + mask interface adapters was returned to fisher & paykel healthcare (f&p) in new zealand, where it was visually inspected.Result: visual inspection of the returned opt980e optiflow + mask interface adapter revealed that the tubing was stretched and detached at the 3-way connector.The customer later reported that the provided lanyard of the interface was not used.Conclusion: we were unable to determine the cause of the reported fault.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is disposed of.The subject interface would have met the specification at the time of production.The user instructions which accompany the opt980e optiflow + mask interface adapter show in pictorial format the correct placement and fitting of the cannula and also warn: - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "do not crush or stretch tube." - "do not soak, wash, sterilize, or reuse this product.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + MASK INTERFACE ADAPTER
Common Device Name
BZA
MDR Report Key11240994
Report Number9611451-2021-00102
Device Sequence Number3726538
Product Code BZA
UDI-Device Identifier09420012436948
UDI-Public(01)09420012436948(10)2101225661(11)200806
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K162553
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT980E
Device Catalogue NumberOPT980
Device Lot Number2101225661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2021
Initial Date Received by Manufacturer 12/29/2020
Supplement Date Received by Manufacturer03/07/2021
Initial Report FDA Received Date01/27/2021
Supplement Report FDA Received Date04/05/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
F&P 900PT290E HUMIDIFICATION CHAMBER; F&P 900PT560 HEATED BREATHING TUBE; F&P PT101 AIRVO 2 HUMIDIFIER
-
-