QUICK-SET PARADIGM; UNO QUICK-SET 80/9 PCC MECA Back to Search Results
Model Number MMT-386
Medical Device Problem Codes Detachment of Device or Device Component (2907); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in the united states.On (b)(6) 2020, it was reported that, the infusion set's cord/tubing was broken 7 inches from where it connects closer to the site, while she was sitting in the couch.She put it this morning and when she looked down, it was observed that both the pieces landed on the couch.The site location was patient's left abdomen and the pump was in her left pocket.Moreover, the infusion had been used for one day.Reportedly, the infusions were not stored or used in a place where they might have been exposed to extreme temperatures and humidity.Further, there was stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.Currently, patient's blood glucose level was 130 mg/dl.No further information available.
 
Event or Problem Description
On 22-apr-2021: follow up information was submitted to update awareness date and result of complaint investigation of the returned used device (1 tubing) showed that the tubing was returned damage (cut).Based on the result: medical device problem code, component code, type of investigation code, investigation findings code and investigation conclusions code were updated.Unomedical reference number (b)(4).Event occurred in the united states.On (b)(6) 2021, it was reported that, the infusion set's cord/tubing was broken 7 inches from where it connects closer to the site, while she was sitting in the couch.She put it this morning and when she looked down, it was observed that both the pieces landed on the couch.The site location was patient's left abdomen and the pump was in her left pocket.Moreover, the infusion had been used for one day.Reportedly, the infusions were not stored or used in a place where they might have been exposed to extreme temperatures and humidity.Further, there was stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.Currently, patient's blood glucose level was 130 mg/dl.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICK-SET PARADIGM
Common Device Name
UNO QUICK-SET 80/9 PCC MECA
MDR Report Key11241534
Report Number3003442380-2021-00118
Device Sequence Number14341687
Product Code FPA
UDI-Device Identifier05705244006577
UDI-Public05705244006577
Combination Product (Y/N)Y
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date02/01/2023
Device Model NumberMMT-386
Device Lot Number5310556
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2021
Supplement Date Received by Manufacturer03/23/2021
Initial Report FDA Received Date01/27/2021
Supplement Report FDA Received Date04/22/2021
Patient Sequence Number1
-
-