C.R. BARD, INC. (BASD) -3006260740 M.R.I. LOW-PROFILE IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 6.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 0603880C |
Device Problems
Fracture (1260); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The catalog number identified has not been cleared in the us but is similar to the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products is identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation; however, photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 10/2022).
|
|
Event Description
|
It was reported that during a port placement procedure through right internal jugular vein, the introducer needle was allegedly found damaged.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a complaint history review was performed.This is the fifth complaint reported for this product/lot number combination.The device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for these reports.Investigation summary: the sample was not returned for evaluation, two electronic photos were provided for review.The investigation is inconclusive for the reported introducer needle hub/syringe damage as the provided photo is not clear enough to confirm the issue.A definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products that are cleared in the us.The pro code and 510k number for the m.R.I.Low-profile implantable port, hickman single-lumen, 6.6f products is identified in d2 and g4.H10: d4 (expiry date: 10/2022).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
|
|
Event Description
|
It was reported that during a port placement procedure through right internal jugular vein, the connection between the syringe and the introducer needle was damaged.There was no reported patient injury.
|
|