Model Number 1961-92-011
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Synovitis (2094); Discomfort (2330); Joint Laxity (4526); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 12/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision of right total knee to address extensive scarring of the extensor mechanism to the distal femur : adhesions, date of implantation: (b)(6) 2018, date of revision: (b)(6) 2020, (right knee).Treatment: tibial insert was revised.Product details: catalog: 960102, lot: 8749379, description: oval dome patella 3 peg sz 38; catalog: 196192011, lot: 8250488, description: aox cvd rp tibial insert sz 2 10mm; catalog: 129433120, lot: 8645520, description: mbt cemented keel sz 2; catalog: 960012, lot: d16053222, description: sigma fem c/ret cemented rt sz 2; catalog: 3332040, lot: 8716090, description: smart set bone cement cmw3; catalog: 3332040, lot: 8716090, description: smart set bone cement cmw3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
The primary operative note, multiple post-operative notes, and revision operative note were reviewed.In additional to the previously alleged patient harms, the patient was also experiencing discomfort and instability prior to revision.The revision operative note also indicated a small amount of erosion just proximal to the femoral flange discovered.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG RP AOX CRV INS SZ2 10
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11243020
Report Number1818910-2021-01928
Device Sequence Number3806079
Product Code NJL
UDI-Device Identifier10603295076087
UDI-Public10603295076087
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1961-92-011
Device Catalogue Number196192011
Device Lot Number8250488
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/15/2021
Supplement Date Received by Manufacturer01/25/2021
01/25/2021
02/19/2021
Initial Report FDA Received Date01/28/2021
Supplement Report FDA Received Date02/10/2021
02/10/2021
02/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY CMW 3 40G.; DEPUY CMW 3 40G.; MBT CEM KEEL TIB TRAY SZ2.; PFC*SIGMA C/R NPOR FEM RT SZ 2.; PFC*SIGMA/OV/DOME PAT 3PEG,38.; SIG RP AOX CRV INS SZ2 10.; DEPUY CMW 3 40G; DEPUY CMW 3 40G; MBT CEM KEEL TIB TRAY SZ2; PFC*SIGMA C/R NPOR FEM RT SZ 2; PFC*SIGMA/OV/DOME PAT 3PEG,38; SIG RP AOX CRV INS SZ2 10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient Weight59
-
-