| Model Number |
G09691 |
| Medical Device Problem Codes |
Device Contamination with Chemical or Other Material (2944); Packaging Problem (3007)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Occupation: other non-healthcare professional: inventory analyst.Pma/510k #: preamendment.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
As reported, when receiving a performer introducer, the customer noticed a white powdery substance within the sealed package.No adverse effects were reported due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Corrected information: b5.It was initially reported that foreign matter was found in the package of two devices from the same lot.The original mdr did not specify a quantity of two.H3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Foreign matter was found in the package of two devices from the same lot.
|
| |
|
Event or Problem Description
|
|
No new patient or event information to report.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Description of event: as reported, there were small white particles within the packaging the performer introducer device.It was also noted that the device was packaged differently than normal.The dilator was not in the sheath but was packaged separately within the set.Investigation ¿ evaluation: a visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record, drawings, the instructions for use, manufacturing instructions, quality control data, and trends.The complainant returned two rcfw-16.0p-38-30-rb unused to cook for investigation.Physical examination of the returned device showed: two sealed devices received.Inspection found foreign matter on the dilator hubs and in sealed pouch.Supplier investigation: customer could not provide a sample of the defect to oliver, therefore sample was sent to the customers lab for ftir testing.The report showed the white particles to be the adhesive applied to the tyvek.This is parent material and would not be considered foreign material.Product had been placed into the pouch prior to the adhesive particles being found.These particles found are likely due to the object inserted into the pouch scraping against the coated tyvek.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: "how supplied supplied sterilized by ethylene oxide gas in peel-open packages.Intended for one-time use.Sterile if package is unopened and undamaged.Do not use the product if there is doubt as to whether the product is sterile.Store in a dark, dry , cool place.Avoid extended exposure to light.Upon removal from package, inspect the product to ensure no damage has occurred." a review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Based on the provided evidence, supplier investigation and the completed investigation, cook has concluded transportation contributed to this incident.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|