• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND Back to Search Results
Model Number 1000354
Device Problems Loose or Intermittent Connection (1371); Use of Device Problem (1670)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/30/2020
Event Type  malfunction  
Manufacturer Narrative
Per tandem¿s user guide: ¿check that your connection between the cartridge tubing and the infusion set tubing is tight and secure. ¿ no product was returned for evaluation. Should new relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the luer lock disconnected. Customer's blood glucose was 455 mg/dl. An infusion set change was performed to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameT:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of DeviceAUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key11244126
MDR Text Key229137729
Report Number3013756811-2021-09640
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1000354
Device Catalogue Number1006379
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received01/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 01/28/2021 Patient Sequence Number: 1
Treatment
INSULIN TYPE: NOVOLOG / NOVORAPID
-
-