The manufacturer received information alleging a ventilator inoperative condition occurred.The device was not in patient use.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device's machine flow sensor was replaced to address the issue.During the evaluation of the device at the manufacturer's service center, a "service required" code was found in the ventilator's downloaded error log.The device's system board was replaced to address the issue.
|