• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 52 DISCOVERY RAD-5 PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 52 DISCOVERY RAD-5 PULSE OXIMETER Back to Search Results
Model Number 26818
Medical Device Problem Code Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was evaluated.External visual inspection found no damage.The device was unable to power on during functional testing.Internal inspection found the regulator of the system board had failed.Heating was detected while measuring the u7 component of the system board causing the warming of the unit.The customer complaint was duplicated in regard to not powering on, the device does not store log files so the allegation of shutting down during use could not be confirmed.A service history record review reveals that this unit was in the field for over one (1) year with one previously reported issue related to this reported event.
 
Event or Problem Description
The customer reported the device turned off over night and felt warmer than usual.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-5 PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
MDR Report Key11244783
Report Number3011353843-2021-00015
Device Sequence Number14341694
Product Code DQA
UDI-Device Identifier00843997005517
UDI-Public00843997005517
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K033296
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number26818
Device Catalogue Number9196
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2020
Initial Report FDA Received Date01/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-