| Model Number |
7209494 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a shoulder surgery, the gripper of the "suture manipulator" did not fully closed.The procedure was successfully completed without delay through a change in the surgical technique.No further complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of the instructions for use found:¿ read these instructions completely prior to use.As with any surgical instrument, careful attention should be exercised to ensure that excessive force is not placed on the instrument.Excessive force can result in instrument failure.¿ a review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A visual inspection revealed the gripper tip has a gap between the pins.A functional evaluation revealed the device requires additional force to close the gap between the distal tip pins.The complaint was confirmed, the root cause was associated with a component failure.Factors that could have contributed to the reported event include wear from use.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|