• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS LIAT SYSTEM; RESPIRATORY VIRUS PANEL NUCLEIC ACID ASSAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS LIAT SYSTEM; RESPIRATORY VIRUS PANEL NUCLEIC ACID ASSAY SYSTEM Back to Search Results
Catalog Number 07341920190
Medical Device Problem Codes False Positive Result (1227); Unable to Obtain Readings (1516); Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
Health Effect - Clinical Codes Anxiety (2328); Unspecified Respiratory Problem (4464); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is ongoing.A follow-up report will be filed upon the completion of the investigation.The test used on the cobas liat system is the cobas sars-cov-2 & influenza a/b test for use on the cobas liat system (product code: qjr).The product catalog number for the test is 09211101190 and the udi is (b)(4).(b)(4).
 
Event or Problem Description
A customer in the us initially reported the generation of invalid results and one discrepant result when using the cobas liat sars-cov-2/flu test on the cobas liat system.The alleged sample generated a sars-cov-2 not detected, flu a not detected, flu b detected result in the original test.Repeat test generated not detected results for all targets.During the review of the data provided for investigation, it was identified that 2 additional samples had false positive results for sars-cov-2 and flu b.No patient information was provided and no harm was indicated.Three mdrs, one per each alleged result will be submitted.
 
Additional Manufacturer Narrative
Review of the run data showed the potential false positive results were caused by abnormal baseline/photometer readings.Inspection of the returned cobas liat analyzer found evidence of tube leakage and debris was found on the optical window and at the edges of the window.Additionally, review of the data identified one more run with a potential false positive sars-cov-2 result related to abnormal baseline/photometer readings.As such, one additional mdr will be filed.Roche received complaints alleging invalid and/or false positive results with the cobas® sars-cov-2 & influenza a/b test for use on the cobas® liat® system for one or more targets (sars-cov-2, influenza a, influenza b).When reviewing the customer-provided data associated with the reported invalid and false positive results, abnormal pcr curves were observed.Per the on-going investigation, several potential causes for the abnormal pcr growth curves leading to invalids and false positives have been identified.These include tube leaks, abnormal pcr steps, and loose thermal sensor wiring.Overall across the installed base, these issues from product use may occur sporadically.For invalid or false positive influenza results, adverse health consequences are not likely.For invalid sars-cov-2, adverse health consequences are not likely since detectability is high and testing can be performed on alternative platforms.For erroneous positive sars-cov-2 results, there is the possibility of adverse health consequences in high risk individuals.As stated in the instructions for use, clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status.A cobas liat software update has been launched to better identify the thermal sensor errors.A new cobas® sars-cov-2 & influenza a/b script to better detect abnormal pcr curves will be made available in due course.Consignees have been notified.G1 - mfg site information were updated to (b)(4).(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS LIAT SYSTEM
Common Device Name
RESPIRATORY VIRUS PANEL NUCLEIC ACID ASSAY SYSTEM
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
MDR Report Key11247389
Report Number2243471-2021-00188
Device Sequence Number4108029
Product Code OCC
UDI-Device Identifier07613336100097
UDI-Public07613336100097
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K111387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07341920190
Initial Date Received by Manufacturer 12/29/2020
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date01/28/2021
Supplement Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2243471-03-17-2021-001-C
Patient Sequence Number1
-
-