It was decided to remove patella and replace with larger size and replaced existing poly with new one of the same size.No delay reported.The affected complaint device, used in treatment, was returned and evaluated.A visual inspection of the returned device shows cement debris on the surface of the part.Damage is observed on the surface of the device.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode.Batch number is not provided, thus a manufacturing records could not be performed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.A clinical evaluation noted that per email correspondence, the requested medical/surgical records and x-rays are not available, and no additional information has been provided.Therefore, the patient impact beyond the revision cannot be determined.Due to the limited information provided, no further clinical assessment is warranted.No further medical assessment can be rendered at this time.Some potential causes of the reported event could include but are not limited to size of device or surgical technique used.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|