| Model Number |
C8702 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: lap colon.Event description: transparant seal came of and fell into patient.Patient status: no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection of the event unit confirmed that the shield, an internal plastic component of the seal, had separated from the seal.Engineering also observed that the septum, an internal rubber component of the seal, had been torn.Based on the condition of the returned unit and the description of the event, it is likely that the reported event was caused by non-axial insertion or removal of asymmetrical instruments through the trocar.Applied medical¿s instructions for use (ifu) states that, "extra care should be used when inserting angular and asymmetrical instruments, such as 'j' hooks and clip appliers.All instruments should be centered axially when inserted through the seal to prevent tearing.".
|
| |
|
Event or Problem Description
|
|
Procedure performed: lap colon.Event description:
transparant seal came of and fell into patient.Patient status: no patient injury.
|
| |
|