• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG PLUG F/TIBIAL PLATEAU; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG PLUG F/TIBIAL PLATEAU; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number NN260P
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Implant Pain (4561); Swelling/ Edema (4577)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for the reported lot number; the device was found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Event or Problem Description
It was reported to aesculap inc.That an vega knee implant system, comprised of the following components, vega system ps femur f5n r (part #nx031z) (ref.Associated mdr id 2916714-2020-00624), vega system patella 3-peg size p3 (part # nx043) (ref.Associated mdr id 2916714-2020-00625), vega system ps tibia plateau t1, cemented implant (part # nx051z) (ref.Associated mdr id 2916714-2020-00626), vega system ps gliding surface implant 14mm t1/t1+ (part # nx112) (ref.Associated mdr id 2916714-2020-00627), and a vega system peek plug t1-t5 (part # nn260p), was implanted during a primary surgery performed on (b)(6) 2013.According to the complainant, following the primary surgery, the patient experienced pain and swelling in in the right knee, which restricted routine daily activities.The patient underwent a revision surgery performed on (b)(6) 2018.The procedure was completed using another manufacturer's implant.Intraoperative findings during replacement surgery revealed loosening of the tibial and femoral components.No components were available to be returned to the manufacturer for evaluation.No known adverse patient effects occurred as a result of the revision surgery.Additional information has been requested, but has not been made available.The adverse event / malfunction is filed under aag reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUG F/TIBIAL PLATEAU
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key11248052
Report Number2916714-2020-00628
Device Sequence Number3533103
Product Code JWH
UDI-Device Identifier04046963711320
UDI-Public4046963711320
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/28/2021
Report Date (Section F) 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2018
Device Model NumberNN260P
Device Catalogue NumberNN260P
Device Lot Number51950754
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/08/2019
Device Age91 MO
Event Location Hospital
Date Report to Manufacturer04/08/2019
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date01/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/28/2021
-
-