Model Number 71364012
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Case: (b)(4).
 
Event or Problem Description
It was reported that in a total hip replacement the anteversion handle was noticed broke.It is unknown when the event occurred.Smith and nephew backup device was available.No injury or delay reported.
 
Additional Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured, and the fractured piece was returned, rendering the device inoperable.The device was manufactured in 2016 and shows signs of extensive use.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTEVERSION HANDLE
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11248234
Report Number1020279-2021-00742
Device Sequence Number3582664
Product Code JDG
UDI-Device Identifier03596010199287
UDI-Public03596010199287
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71364012
Device Catalogue Number71364012
Device Lot Number16CSC0005
Initial Date Received by Manufacturer 01/08/2021
Supplement Date Received by Manufacturer02/02/2021
Initial Report FDA Received Date01/28/2021
Supplement Report FDA Received Date02/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-