Model Number N/A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported an unknown patient required an ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter for bile duct drainage.Immediately following placement, when the physician applied negative pressure, bile leaked between the mac-lock hub and catheter.The device was removed and replaced with a similar device to complete the procedure successfully.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation.(b)(6) informed cook that an ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter from lot 13482348 leaked during placement.The catheter was placed in the bile duct percutaneously.When the physician applied negative pressure right after placement, fluid leaked between the mac-lock hub and catheter.Therefore, another drainage catheter was used instead.There have been no adverse effects to the patient reported.A review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control of the device, as well as a visual inspection, functional test, and dimensional verification, were conducted during the investigation.The customer returned one used 8.5fr ult catheter.There is biological matter throughout and no visible damage.A leak test confirmed leakage where the catheter enters the connector cap.The distance between the cap and the hub was measured and determined to be within specification.Additionally, a document based investigation evaluation was performed.The risk specifications covering mac-loc drainage catheters includes hub separation as a potential failure mode.The identified risk controls include the manufacturing quality control checks and process validation.The technical files covering mac-loc drainage catheters indicate that the risks associated with these devices are acceptable when weighed against the benefits.Based on the review of current documentation, cook has concluded that inspection activities are currently in place to prevent the release of non-conforming product related to the reported failure mode.The ifu supplied with the mac-loc drainage catheters instruct that the product should be inspected prior to use to ensure no damage has occurred.A review of the device history record (dhr) was also conducted as a part of the investigation to check for failure related nonconformances and additional complaints.Lot 13482348 did not have related nonconformances.There are no additional complaints on this lot.There is no evidence of nonconforming material in house or in the field.There is no evidence of manufacturing deficiency, therefore, the cause of this event is component failure.The appropriate personnel have been notified.Per the quality engineering risk assessment no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE DAWSON-MUELLER MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key11248306
Report Number1820334-2021-00199
Device Sequence Number3606712
Product Code GBO
UDI-Device Identifier00827002137660
UDI-Public(01)00827002137660(17)231009(10)13482348
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/09/2023
Device Model NumberN/A
Device Catalogue NumberULT8.5-38-25-P-5S-CLDM-WF-HC
Device Lot Number13482348
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2021
Initial Date Received by Manufacturer 01/21/2021
Supplement Date Received by Manufacturer05/19/2021
Initial Report FDA Received Date01/28/2021
Supplement Report FDA Received Date06/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-