TENEX HEALTH, INC. TENEX HEALTH TX SYSTEM - WITH TXB MICROTIP; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 554-3003-001
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 12/30/2020
Type of Reportable Event Serious Injury
Event or Problem Description
User ran out of saline irrigation during a procedure due to using a smaller than specified size saline bag (250cc instead of 1000cc).They reported the device was overheating as a result, and the patient developed a superficial skin burn at the treatment site.The burn started as a blister and was healing at 10 days post-op.It was reported that the wound was fully closed and the patient was pain-free after approximately 14 days.Patient may be under observation for 60 days.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENEX HEALTH TX SYSTEM - WITH TXB MICROTIP
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630 8123
MDR Report Key11249071
Report Number1000135560-2021-00012
Device Sequence Number3563033
Product Code LFL
UDI-Device Identifier00857156006093
UDI-Public00857156006093
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K181367
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number554-3003-001
Device Catalogue Number554-3003-001
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2020
Initial Report FDA Received Date01/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age49 YR
-
-