| Model Number |
0030-4864 |
| Medical Device Problem Code |
Computer System Security Problem (2899)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/31/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The system was evaluated by a field service engineer.The field service found no issues with the unit.A field service checklist was performed.A record review was performed.A product deficiency review was performed and there is no product deficiency identified.A document, service history, and trending was reviewed.There is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.A labeling review was conducted; the operator manual for the system was reviewed and found to include adequate instructions for use, warnings and operational errors.The review of the device history record (dhr) for femtosecond laser system showed that there were no issues or non-conformities.The system and its components met all specifications prior to being released.Manufacturing has been ruled out as a potential cause for the reported issue.Based on the investigation results, no corrective action has been issued.Based on the investigation results there is no indication of a product quality deficiency.Johnson & johnson surgical vision will continue to monitor this type of complaints all pertinent information available to johnson & johnson surgical vision, inc has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported that their treatment was interrupted with 12 seconds left in a photorefractive keratectomy (prk) treatment.They were unable to boost the laser, and complete the treatment.Patient was ok and no patient injury reported.
|
| |
|
Search Alerts/Recalls
|