| Catalog Number |
CM-9145 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination of the returned product identified obvious signs of use.The anchor is fractured near the tip and all pieces appear to be present.The anchor is still attached to the inserter and could be removed from the inserter.One of the prongs of the inserter is bent and makes removing the anchor difficult but possible.The handle has no obvious signs of damage and the knob and switch are functional.Medical records were not provided.Dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately three (3) weeks ago during a rotator cuff repair, the tip of the anchor broke while the surgeon was advancing the awl.Patient did not experience any harm as a result of the anchor fracturing.Attempts have been made and no further information has been provided.
|
| |
|