| Model Number |
313.843 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Without a lot number the device history records review could not be completed.Product was not returned.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is synthes sales representative.510k: premarket submission number not available/not released.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device was used for treatment, not diagnosis.If additional information is made available, the investigation will be updated as applicable.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, while inspecting the instruments after going through the decontamination process, it was discovered that some instruments were damaged.The star drive screwdriver shaft is bent and wobbles in the screwdriver handle, wire guide sleeve won't fully insert into the blade guide, blade/screw guide sleeve will not let the blade inserter engage the rifling properly, tube part of depth gauge for screws is bent and will not let the screw down piece engage the threads, the insertion guide was bent and the sleeves were unable to be inserted full.However, it was set aside and inadvertently placed in the sharps container by the staff and was unable to be retrieved.There was no patient involvement with any of these items.This complaint involves five (5) devices.Concomitant products: depth gauge for screws ø 2.7 to 4.0 mm, measuring range up to 60 mm, 3.2mm wire guide sleeve.Blade/screw guide sleeve.Depth gauge for screws ø 2.7 to 4.0 mm, insertion guide.This report is for one (1) stardrive screwdriver shaft sd6/self-retaining/2.0mm.This report is 1 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9 h3, h4, h6: device history lot.Part # 313.843.Synthes lot number: h155276.Supplier lot number: h155276.Release to warehouse date: 27.Dec.2016.Manufactured by: monument.No ncr¿s were generated during production.Device history batch , null device history review review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Investigation summary: background: it was reported that on an unknown date, while inspecting the instruments after going through the decontamination process it was noticed that some instruments were damaged.The star-drive screwdriver shaft is bent and wobbles in the screwdriver handle, wire guide sleeve won't fully insert into the blade guide, blade/screw guide sleeve will not let the blade inserter engage the rifling properly, tube part of depth gauge for screws is bent and will not let the screw down piece engage the threads, the insertion guide was bent and the sleeves were unable to be inserted full.However, it was set aside and inadvertently placed in the sharps container by the staff and was unable to be retrieved.There was no patient involvement with any of these items.This complaint involves five (5) devices.H6: investigation flow: damage visual inspection: the stardrive screwdriver shaft sd6/self-retaining/2.0mm (part #: 313.843, lot #: h155276) was received at us cq.Upon visual inspection, the device was bent mid-shaft.Document/specification review: 313_842 rev k (current) and rev h (manufactured) were reviewed.No design issues or discrepancies were identified.Complaint was confirmed.Investigation conclusion: the overall complaint is confirmed as the device was bent at mid-shaft.However the "unable to assemble" reported condition could not be confirmed as no test could be performed.No definitive root cause could be determined based on the provided information but it is likely that the device experienced external forces greater than what the device was validated to withheld.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventive action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|