| Model Number |
SPL-T |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation confirmed the user report, as the transducer plug was damaged with the black end cover detached.The review of the device history record no abnormalities or anomalies identified during production.The device met all specifications upon release.The ifu contains the following statements: "this product is a precision device; handle it with care.Avoid rough or violent handling, which may cause equipment damage," and "take special care to ensure all cables are undamaged." olympus personnel have concluded the observed failure was most likely caused by user mishandling.Olympus will continue to monitor the field performance of this device.
|
| |
|
Event or Problem Description
|
|
As reported, the section of the shockpulse lithotripsy transducer that attaches to the monitor was noted to be broken during device maintenance.There was not reported patient involvement from this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
|
| |
|