• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH MOBILETT MIRA MAX; MOBILE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH MOBILETT MIRA MAX; MOBILE X-RAY SYSTEM Back to Search Results
Model Number 10594738
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is on-going.In the unlikely event of both mounting brackets breaking off and the collimator becoming completely loose, it would likely be held by the cables.It is described in the operator manual to perform a daily check for damaged or loose parts and have the system repaired if problems are found.A supplemental report will be submitted if additional information becomes available.Internal id (b)(4).
 
Event or Problem Description
It was reported that one of two collimator cover brackets of the mobilett mira max snapped off.The structure of the unit remained intact; only slight wobbling was visible.No risk of falling parts was reported.However, the entire collimator to avoid any potential risks.There are no injuries attributed to this event.
 
Additional Manufacturer Narrative
The issue was investigated in detail and based on information from similar cases.The collimator cover is fixated by four screws on two holders, which are part of the collimator.The mounting brackets are produced with steel sheet metal bent to 90° with the addition of four threaded inserts.The housing is mounted to these four threaded inserts.Upon investigation, it was discovered that the sheet metal had been bent two times in opposite directions, thus weakening the piece.As a result, the sub supplier was changed.Furthermore, the design was modified to ensure additional strength of the piece; visual checks were implemented to ensure the piece was neither broken nor bent two times.These mentioned improvements on the brackets have been already introduced by the supplier for collimators used for mobilett elara units, starting with serial number (b)(6) (in combination with collimator serial number (b)(6)).This collimator is used on two system types with different part numbers.Mobilett mira max (10519819), mobilett elara max (11364066).According to the user manual (xpr8-270g.620.03.01.02, page 123), the user must perform a daily check of function and security.If damaged or loose parts are found, the system may not be used until repair was performed.The described issue was resolved at the concerned customer site by siemens local service technician replacing the affected part.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBILETT MIRA MAX
Common Device Name
MOBILE X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
henkesstrasse 127
erlangen, 91052
GM  91052
MDR Report Key11252578
Report Number3002808157-2021-65370
Device Sequence Number3198456
Product Code IZL
UDI-Device Identifier04056869009131
UDI-Public04056869009131
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111912
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10594738
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/25/2021
Supplement Date Received by Manufacturer04/13/2021
Initial Report FDA Received Date01/29/2021
Supplement Report FDA Received Date04/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-