MICROVENTION, INC. HYDROFILL 10 ADVANCED; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 7310-0615-A2
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A search for non-conformance associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
 
Event or Problem Description
It was reported that during a coil embolization procedure, the embolization coil detached in the microcatheter.The coil was removed from the patient without incident.There was no reported patient injury or additional intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDROFILL 10 ADVANCED
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
MDR Report Key11255516
Report Number2032493-2021-00025
Device Sequence Number3477376
Product Code HCG
UDI-Device Identifier00816777023407
UDI-Public(01)00816777023407(11)200131(17)241231(10)2001311W5
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K113457
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number7310-0615-A2
Device Lot Number2001311W5
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/31/2020
Initial Report FDA Received Date01/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age42 YR
-
-