| Model Number |
ESR9222400 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.Coloplast has not been provided any corroborating evidence to verify the information contained in this report.
|
| |
|
Event or Problem Description
|
|
According to the available information, though not verified, tubing break.Device removed and replaced.Additional information received 01/05/2020: the break was between the pump and left cylinder.Approximately 1cm from the protective collar of the tubing/pump junction.The lot # was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.
|
| |
|