• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN TAYLOR SPATIAL FRAME EXFIX DEV; PIN, FIXATION, THREADED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN TAYLOR SPATIAL FRAME EXFIX DEV; PIN, FIXATION, THREADED Back to Search Results
Catalog Number UNKN02300100
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Swelling/ Edema (4577)
Event Date 08/05/2011
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that in a clinical observation of "pseudoaneurysm of the peroneal artery: an unusual complication of open docking site procedure in bone transport with taylor spatial frame" that the taylor spatial frame (smith & nephew) was applied to 1 patients and presented a readmission due to pain, swelling and wound bleeding.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the data presented in the aged article, "pseudoaneurysm of the peroneal artery: an unusual complication of open docking site procedure in bone transport with taylor spatial frame" did not provide insight or relevance to current clinical outcomes for the product/device.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation of the report, readmission due to pain, swelling and wound bleeding could not be performed.The images provided in the article have been interpreted within the text; therefore, no further analysis of the images is required.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction or loss of sterility.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN TAYLOR SPATIAL FRAME EXFIX DEV
Type of Device
PIN, FIXATION, THREADED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11256338
MDR Text Key229532930
Report Number1020279-2021-00793
Device Sequence Number1
Product Code JDW
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKN02300100
Device Lot NumberUNKNOWN
Initial Date Manufacturer Received 01/11/2021
Initial Date FDA Received01/30/2021
Supplement Dates Manufacturer Received07/08/2021
Supplement Dates FDA Received07/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-