| Model Number |
MMT-551NAS |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Pump received unable to perform the displacement test due to broken/detached end cap.Pump received with cracked battery tube threads, partial broken reservoir tube lip, cracked reservoir tube window and cracked case reservoir tube window corner.The p-cap locks into the reservoir compartment properly noted.Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the bottom of insulin pump was cracked.No harm requiring medical intervention was reported.The insulin pump was returned for analysis.
|
| |
|