• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC GYN OPERATIVE 12 DEGREE 2.7MM LONG AUTOCLAVABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC GYN OPERATIVE 12 DEGREE 2.7MM LONG AUTOCLAVABLE Back to Search Results
Model Number G27L-12A
Medical Device Problem Codes Break (1069); Defective Device (2588)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was evaluated.Inspection of the device confirmed the reported issue.Device was found with fractured internal lens.Investigation is ongoing therefore the root cause cannot be determined at this time.If additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
It was reported that the device was found with fractured internal lens.The issue occurred during reprocessing.There was no patient involvement on this event reported.No user injury reported.
 
Additional Manufacturer Narrative
This supplemental report is to inform that upon further review, this is not a reportable malfunction.Per the legal manufacturer, there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYN OPERATIVE 12 DEGREE 2.7MM LONG AUTOCLAVABLE
Common Device Name
GYN OPERATIVE 12 DEGREE 2.7MM LONG AUTOCLAVABLE
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key11261542
Report Number1519132-2021-00003
Device Sequence Number728527
Product Code FHO
UDI-Device Identifier00821925008915
UDI-Public00821925008915
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K980972
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG27L-12A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/08/2021
Supplement Date Received by Manufacturer02/09/2021
Initial Report FDA Received Date02/01/2021
Supplement Report FDA Received Date02/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-