| Model Number |
81002 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation: the service representative was able to confirm the reported condition.The iv pole lock was adjusted.Iv pole was tested.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Root cause: the root cause of the reported incident was the screw that pushes the iv pole release button needed to be tightened.
|
| |
|
Event or Problem Description
|
|
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
|
| |
|