| Model Number |
HT70PM-JP-NA |
| Medical Device Problem Codes |
Accessory Incompatible (1004); Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.The service personal(sp) inspected the unit and confirmed the reported issue.Sp replaced the control board and filter assembly to address the issue.Sp also performed the preventive maintenance (pm) and replaced the power pack battery, internal battery cooling fan and oxygen sensor.It was reported that during servicing, the inlet filter for this ht70 ventilator was damaged.Additionally, it was reported that there was no alarm sound generated during shutdown.The reported issue was/could not be confirmed.The most likely cause was isolated in the field to a potential fault in replaced components filter assembly and control board.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during servicing, the inlet filter for this ht70 ventilator was damaged.Additionally, it was reported that there was no alarm sound generated during shutdown.The ventilator was not in use on a patient at the time of the reported event.
|
| |
|