| Model Number |
400SMTXSFT0202 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-00202, 3005168196-2021-00203, 3005168196-2021-00205.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the middle cerebral artery (mca) using penumbra smart coils (smart coils).During the procedure, four smart coils would not advance at all within their introducer sheaths; therefore, the four smart coils were removed.The procedure was completed using additional smart coils.There was no report of an adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|