• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTXSFT1H03
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2021-00202, 3005168196-2021-00203, 3005168196-2021-00204.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the middle cerebral artery (mca) using penumbra smart coils (smart coils).During the procedure, four smart coils would not advance at all within their introducer sheaths; therefore, the four smart coils were removed.The procedure was completed using additional smart coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned first smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced while advancing the device.Forcefully advancing the device against this resistance may have contributed to the kink and fracture which were observed near the pusher assembly mid-joint.Further evaluation of the device revealed an additional pusher assembly kink and a detached embolization coil.This damage was likely incidental to the complaint.Evaluation of the returned second smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If this occurs, resistance will likely be experienced while advancing the device.Forcefully advancing the device against this resistance may have contributed to the fractures which were observed near the pusher assembly mid-joint.Further evaluation of the device revealed a pusher assembly kink and a detached embolization coil.This damage was likely incidental to the complaint.Evaluation of the returned smart coil revealed offset coil winds along the length of its embolization coil and the embolization coil was returned partially advanced distal to the introducer sheath.Based on the returned condition, the reported complaint that the coil could not be advanced at all within its introducer sheath could not be confirmed.If the introducer sheath is not properly aligned within the hub of the microcatheter prior to advancement, resistance maybe encountered and damages such as offset coil winds may occur.During functional testing, the smart coil was unable to advance from its returned position due to the dried blood inside its introducer sheath.The dried blood inside the introducer sheath was likely incidental to the complaint.Further evaluation of the device revealed kinks along the pusher assembly.This damage was likely incidental to the complaint.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 1.3005168196-2021-00202, 2.3005168196-2021-00203, 3.3005168196-2021-00204.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11264413
Report Number3005168196-2021-00205
Device Sequence Number1219249
Product Code HCG
UDI-Device Identifier00814548015231
UDI-Public00814548015231
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/06/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/27/2022
Device Model Number400SMTXSFT1H03
Device Catalogue Number400SMTXSFT1H03
Device Lot NumberF78888
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/06/2021
Supplement Date Received by Manufacturer02/25/2021
Initial Report FDA Received Date02/02/2021
Supplement Report FDA Received Date03/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
Date Report Sent to FDA01/01/2005
-
-