Model Number 1829500
Device Problems Malposition of Device (2616); Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/07/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history record could not be performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.
 
Event Description
It was reported that approximately three months post port placement, the device allegedly malpositioned.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: the sample was not returned for evaluation, one electronic photo were provided for review.The investigation was confirmed for the reported material protrusion issue as one of the lumen was protruding/dislodging from the port body.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: h6 (method).H11: h6 (device, result and conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
 
Event Description
It was reported that approximately three months post port placement, the device allegedly malpositioned.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11264434
MDR Text Key230752252
Report Number3006260740-2021-00173
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027185
UDI-Public(01)00801741027185
Combination Product (y/n)N
PMA/PMN Number
K090512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1829500
Device Catalogue Number1829500
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received07/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-