C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Model Number 1829500 |
Device Problems
Malposition of Device (2616); Material Protrusion/Extrusion (2979)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was not provided, a review of the device history record could not be performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.
|
|
Event Description
|
It was reported that approximately three months post port placement, the device allegedly malpositioned.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: the sample was not returned for evaluation, one electronic photo were provided for review.The investigation was confirmed for the reported material protrusion issue as one of the lumen was protruding/dislodging from the port body.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: h6 (method).H11: h6 (device, result and conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
|
|
Event Description
|
It was reported that approximately three months post port placement, the device allegedly malpositioned.There was no reported patient injury.
|
|
Search Alerts/Recalls
|
|
|