STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #3 L-CEM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
5510-F-301 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Codes |
Pain (1994); Ambulation or Postural Difficulties (2544)
|
| Date of Event |
01/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left knee was revised due to instability and pain.A cr femur and insert were revised to a ts femur, ps insert, cemented stem and augments.Rep provided primary and revision usage and a picture of the explanted liner, and confirmed that no further information will be released by the hospital or surgeon.
|
| |
|