| Model Number |
204713 |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Osseointegration Problem (3003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/28/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This implant loss suggests that the source of the problem was likely to stem from no osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, overall health or behavior.Proper intended use of the implant is described in its instructions for use.Physical analysis cannot be performed.
|
| |
|
Event or Problem Description
|
|
Bone quality at the time of implant failure type ii.Time of loss in relation to the implantation was before prosthetic restoration.Osseointegration was not achieved.
|
| |
|