|
ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
FF-410535 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
10/23/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified deformation, crease fold, voids in the gel and yellow particles.Weight measurement identified device weight to the specification.Cloudy and voids was observed after autoclave disinfection process.Based on the device analysis, the final assessment is no issues found related with the manufacturing process.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side exchange of textured breast implants due to the patient¿s concern with the product.Additionally, healthcare professional reported left side capsular contracture, baker grade iii.Device was explanted.
|
| |
|
Search Alerts/Recalls
|
|
|