SMITH & NEPHEW, INC. R3 20DEG 36MM TRIAL 54MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Model Number |
71367954 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during surgery, a fin on the outside of the trial insert broke off of the trial, inside the patient.The surgeon did not know whether it happened during seating of trial liner or extraction of trial liner.The liner trial, as well as the full piece of the fin that broke off, was removed.Procedure was completed with the same device, there was no delay.No piece of the liner was left in the patient according to the surgeon: the piece was matched up with the liner and the surgeon was confident that nothing was retained.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device, used in treatment, was returned for evaluation.A visual inspection confirmed the r3 20deg 36mm trial 54mm shows nicks and scratches.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|