BIOTRONIK SE & CO. KG INTICA 7 HF-T QP DF-1 IS4 PROMRI; CRT-D Back to Search Results
Model Number 404629
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Endocarditis (1834)
Date of Event 01/28/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this system was explanted due to endocarditis.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
An endocarditis was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In conclusion, the endocarditis was not device related.
 
Additional Manufacturer Narrative
The method, result, and conclusion codes remain the same as follow up one.An endocarditis was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In conclusion, the endocarditis was not device related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTICA 7 HF-T QP DF-1 IS4 PROMRI
Common Device Name
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key11267338
Report Number1028232-2021-00567
Device Sequence Number19616024
Product Code NIK
UDI-Device Identifier04035479142162
UDI-Public04035479142162
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model Number404629
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2021
Initial Date Received by Manufacturer 01/29/2021
Supplement Date Received by Manufacturer02/02/2021
04/21/2021
Initial Report FDA Received Date02/02/2021
Supplement Report FDA Received Date02/03/2021
04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age55 YR
-
-