| Model Number |
442193 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ plus aerobic/f culture vials false positive results were obtained by the laboratory personnel.A false positive may lead to treatment with a less tolerated antibiotic which can lead to serious adverse patient consequence.There was no report of patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: customer reported a false positive defect.Bd was unable to reproduce customer experience with the bactec product.A false positive response was not observed when retention samples were tested.Batch history records were reviewed, and all testing were within specification for product release.Users are cautioned in the package insert under limitation of the procedure: ¿a gram-stained smear from culture medium may contain small number of non-viable organisms derived from media constituents, staining reagents, immersion oil, glass slide and specimens used for inoculation.There are many factors that can influence the false positive rate, including blood volume, blood cell counts, environmental factors, and media lot to lot variations.Complaint is unconfirmed based on retention samples and batch history record review.No correctives actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigation process.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bactec¿ plus aerobic/f culture vials false positive results were obtained by the laboratory personnel.A false positive may lead to treatment with a less tolerated antibiotic which can lead to serious adverse patient consequence.There was no report of patient impact.
|
| |
|
Search Alerts/Recalls
|