| Model Number |
OCB5014BA |
| Medical Device Problem Codes |
Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
11/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the details of the case provided by the local sales representative and the hospital staff, there is no indication that the flash ostial system device did not perform as intended.The likely root cause for this incident was inflation of the balloon beyond the rated burst volume.In addition to the case details, ostial corporation reviewed the manufacturing documentation and lot release testing associated with the flash ostial system device lot that is suspected to have been used in this procedure.No issues were noted that would have contributed to the reported incident.The instructions for use for the flash ostial system were also reviewed and it was confirmed that vessel dissection is listed as a potential complication associated with the procedure and that the rated burst volume for the balloons are clearly listed.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a tavr procedure and during the procedure it was determined that dilatation of a stent placed in the left main coronary artery was also required.Although not indicated, the physician elected to use the flash ostial system device to post dilate the stent.The outer balloon was inflated beyond the rated burst volume to expand the balloon to a larger diameter than intended per the product labeling.After exceeding the rated burst volume, the balloon ruptured, and an edge dissection was noted distal to the stent.A covered stent was placed to address the dissection and no further clinical sequelae were reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|