| Model Number |
4720 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a breach in the sterile barrier.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: it was reported that the hospital staff noticed a puncture to the front corner of the foil packaging on the new air device.Ahs mdip # 8825.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: 1) severe shipping conditions or 2) user mishandling.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacturing date is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a breach in the sterile barrier.
|
| |
|